- Description
- Specifications
- Setup Software
- Mechanical
- Documents
- Technical FAQ
- Applications
- Accessories
Features
- Reads process signals from ±200 mV to ±600V or ±2 mA to ±5A full scale
- Accuracy ±0.01% of reading ± 2 counts
- Absolute and ratiometric mode for bridges and potentiometers
- Error less than 0.01% of full scale for absolute ranges, less than 0.01% of reading for ratiometric measurements
- All input ranges are user selectable and factory calibrated
- Up to 60 conversions per second, Ideal for peak or valley capture
- Digital span adjust from 0 to ±99,999, zero adjust from -99,999 to +99,999
- Front panel scalable to ±99,999 for use with current shunts
- 1/8 DIN size with bright red or green 0.56" (14.2mm), high LED digits
- Transducer excitation output, 5, 10, 12, or 24 Vdc (isolated)
- Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
- Operating temperature from -40°C to 70°C (-40°F to 158°F)
- Wide choice of Plug-in-Play options:
- 2 or 4 relays, mechanical or solid state, for alarm or control (isolated)
- 1 or 2 Analog output, 4-20 mA, 0-20 mA, 0-10V, or -10V to +10V (isolated)
- Communications: Ethernet, WiFi, USB, RS232, RS485 (isolated)
- Extended DPM allows up to 180 data points for custom curve linearization
and a rate derived from consecutive readings
Laureate™ 1/8 DIN Panel Meter for digital process
are a cost-effective solution for process signals such as 4-20 mA, 0-10V, or 0-5V. Full-scale voltage input ranges from ±200 mV to ±600V and current ranges from ±2 mA to ±5 A are jumper selectable. All ranges are precalibrated at the factory, so that recalibration is not needed when changing ranges or signal conditioners. The 200.00 mV and 2.000 V ranges provide a high input impedance of 1 Gohm to minimize the load on the voltage signal.The panel meters can be set to absolute and ratiometric (or potentiometer follower) mode by making a software selection. In this mode, the meter tracks a ratio of the applied excitation voltage and is unaffected by changes in the excitation voltage. Ratiometric measurements provide an exceptional accuracy of 0.01% of reading ± 2 counts. This capability is used with the panel meter's 5V or 10V excitation output for load cells and Wheatstone bridges, and with the panel meter's 5V excitation output for potentiometers which track wiper position.
All signal conditioner board ranges are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated panel meters. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.
The optional extended Laureate computer board enhances Laureate Panel Meters by displaying rates derived from successive readings and enabling highly accurate custom curve linearization. For example, it can calculate liquid volume or flow rate in a horizontal cylindrical tank using levels from a 4-20 mA transmitter. Setup is straightforward: users input up to 180 data points into a spreadsheet or text file, and the computer calculates spline-fit segments, which are then downloaded to the panel meters for precise operation.
Laureate Panel Meters are easily programmed with Laurel’s free Instrument Setup Software, downloadable from our website and compatible with Windows PCs, requiring a data interface board for setup.
Scaling is from -99,999 to +99,999 (five full digits) with any decimal point to display readings in engineering units, such as PSI. Three scaling methods are user selectable: scale and offset, two-point method, and system-level calibration using actual transducer signals.
High read rate of up to 50 or 60 conversions per second, the Laureate™ Panel Meter uses Concurrent Slope (US Pat. 5,262,780) analog-to-digital conversion to integrate signals over a full power line cycle (50 Hz or 60 Hz). This read rate enables peak and valley capture, real-time computer interfacing, and control applications. Peak and valley values are automatically captured and can be viewed using Laurel’s free Instrument Setup Software (compatible with Windows PCs) or transmitted as serial data.
- An unfiltered selection provides true peak and valley readings and aids in control applications.
- A batch average filter selection averages each 16 conversions.
- An adaptive moving average filter selection provides a choice of 8 time constants from 80 ms to 9.6 seconds. When a significant change in signal level occurs, the filter adapts by briefly switching to the shortest time to follow the change, then reverts back to its selected time constant. An Auto setting selects the time constant selection based on signal noise.
Two tare functions: auto-tare and manual tare. In auto-tare, an input line is grounded by an external pushbutton. In manual tare, the tare value can be entered manually via a control input pushbutton or using Laurel's free Instrument Setup Software.
Peak and valley values are automatically captured. These may be displayed via a front panel pushbutton command or control signal at the rear connector, or be transmitted as serial data.
Two rear panel control Inputs (CMOS/TTL levels, logic 0 = tied to digital ground, logic 1 = open) or dry contacts that can be set to control / activate 14 meter commands.
An (isolated) 5, 10, 12, or 24 Vdc excitation output is standard to power transducers or two-wire transmitters. Ratiometric operation, which automatically compensates for changes in the applied excitation, is jumper selectable for applications, such as bridges, where the signal to be measured is proportional to the excitation level.
DC Voltage
DC Voltage Range | Resolution | Input Resistance | Error at 25°C |
---|---|---|---|
±200.00 | 10 µV | 1 GΩ | 0.01% of FS ± 2 cts |
±2.0000 | 100 µV | 1 GΩ | 0.01% of FS ± 2 cts |
±20.000 | 1 mV | 10 MΩ | 0.01% of FS ± 2 cts |
±200.00 | 10 mV | 10 MΩ | 0.01% of FS ± 2 cts |
±300.00 | 0.1 V | 10 MΩ | 0.01% of FS ± 2 cts |
±600.00* | 0.1 V | 10 MΩ | ± 0.4 V |
* ±600.00V range not ETL certified. | |||
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months. |
DC Current
DC Current Range | Resolution | Input Resistance | Error at 25°C |
---|---|---|---|
±2.0000 mA | 0.1 µA | 100 Ω | 0.01% FS ± 2 cts |
±20.000 mA | 1.0 µA | 10 Ω | 0.01% FS ± 2 cts |
±200.00 mA | 10 µA | 1 Ω | 0.01% FS ± 2 cts |
±5.000 A | 1 mA | 0.01Ω | ± 10 mA |
DC Voltage & Current
Display | |
---|---|
Readout | 5 LED digits, 7-segment, 14.2 mm (.56"), red or green. |
Range | -99999 to 99999 or -99990 to 99990 (count by 10) |
Indicators | Minus sign, 2 red LED lamps |
A-to-D Conversion | |
Technique | Concurrent Slope™ (Pat 5,262,780) |
A-to-D rate | 60/s at 60 Hz, 50/s at 50 Hz |
Output update rate | 56/s at 60 Hz, 47/s at 50 Hz |
Display update rate | 3.5/s at 60 Hz, 3/s at 50 Hz |
Reading Accuracy | |
Error at 25°C | 0.01% of full scale ± 2 counts (except 5A range) for absolute measurements. 0.01% of reading ± 2 counts for ratiometric measurements. |
Span tempco | 0.003% of reading/°C |
Zero tempco | 0.1 count/°C |
Noise Rejection | |
CMR, DC to 60 Hz | 130 dB |
NMR at 50/60 Hz | 90 dB with min filtering |
Maximum Signal | |
Max applied voltage | 600 Vac for 20, 200 and 300 V ranges, 125 Vac for other ranges |
Overcurrent protection |
25x for 2 mA, 8x for 20 mA, 2.5x for 200 mA, 1x for 5 A |
Power | |
Voltage, standard | 85-264 Vac or 90-300 Vdc |
Voltage, optional | 12-32 Vac or 10-48 Vdc |
Frequency | DC or 47-63 Hz |
Power consumption (typical, base meter) | 1.2W @ 120 Vac, 1.5W @ 240 Vac, 1.3W @ 10 Vdc, 1.4W @ 20 Vdc, 1.55W @ 30 Vdc, 1.8W @ 40 Vdc, 2.15W @ 48 Vdc |
Power Isolation | 250V rms working, 2.3 kV rms per 1 min test |
Excitation Output (standard) | |
5 Vdc | 5 Vdc ± 5%, 100 mA (jumper selectable) |
10 Vdc | 10 Vdc ± 5%, 120 mA (jumper selectable) |
12 Vdc | 12 Vdc ± 5%, 100 mA (jumper selectable) |
24 Vdc | 24 Vdc ± 5%, 50 mA (jumper selectable) |
Output Isolation | 50 Vdc from signal ground |
Ratiometric operation | 5 Vdc or 10 Vdc for bridge circuits, 5 Vdc for potentiometers |
Analog Output Board (optional) | |
Output levels | 4-20 mA, 0-20 mA, 0-10V, -10 to +10V (jumper selectable) |
Current compliance | 2 mA at 10V ( > 5 kΩ load) |
Voltage compliance | 12V at 20 mA (< 600 Ω load) |
Scaling | Zero and full scale adjustable from -99999 to +99999 |
Resolution | 16 bits (0.0015% of full scale) |
Step function response | 80 ms to 99% of final value (typ) |
Isolation | 250V rms working, 2.3 kV rms per 1 min test |
Relay Output Boards (optional) | |
Relay types | 2 Form C contact relays or 4 Form A contact relays (NO) 2 or 4 Form A, AC/DC solid state relays (NO) |
Current tatings | 8A at 250 Vac or 24 Vdc for contact relays 120 mA at 140 Vac or 180 Vdc for solid state relays |
Output common | Isolated commons for dual relays or each pair of quad relays |
Isolation | 250V rms working, 2.3 kV rms per 1 min tes |
Communication Boards (optional) | |
Board Selections | RS232, RS485 with dual RJ11 connectors, RS485 with dual RJ45 connectors, USB, High-Speed Ethernet, USB-to-RS485 gateway, High-Speed Ethernet-to-RS485 gateway, WiFi with built-in antenna plus USB & RS485, WiFi with external antenna plus USB & RS485 |
Protocols | Laurel Custom ASCII (serial), Modbus RTU (serial), Modbus TCP (Ethernet or WiFi) |
Digital addresses | 247 (Modbus), 31 (Laurel ASCII), |
Isolation | 250V rms working, 2.3 kV rms per 1 min test |
Signal Connections | |
![]() |
|
Environmental | |
Operating temperature | -40°C to 70°C (-40°F to 158°F) |
Storage temperature. | -40°C to 85°C (-40°F to 185°F) |
Relative humidity | 95% at 40°C, non-condensing |
Protection | NEMA-4X (IP-65) when panel mounted |
Mechanical | |
Enclosure | 1/8 DIN, high impact plastic, UL 94V-0, color: black |
Mounting | 1/8 DIN panel cutout required: 3.622" x 1.772" (92 mm x 45 mm). |
Dimensions | 4.68" x 2.45" x 5.64" (119 mm x 62 mm x 143 mm) (W x H x D) |
Maximum panel thickness | 4.5 mm (0.18") |
Tightening Torque - Connectors | Screw terminal connectors: 5 lb-in (0.56 Nm) |
Tightening Torque - Pawls | Digital Panel Meter Case Pawls: 5 lb-in (0.56 Nm) |
Weight of base meter | 210 g (7.4 oz) typical (DPM, counter, timer, 6-digit remote display) |
Weight of option boards | 30 g (1.0 oz) typical per board (analog output, relay output, communications) |
General | |
Programming Methods | Four front panel buttons or via Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. |
Security | Lockout options include using the front panel buttons, the free Instrument Setup Software, or a hardware jumper. |
Warranty | 3 years parts & labor |
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months. |
Free Instrument Setup Software for Series 2 Laureates
Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.
- When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
- When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.
Download Free Instrument Setup Software
Installation
Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:
- Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
- Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
- Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
- Reboot your computer for the changed UAC setting to take effect.

RJ11-to-DB9 cable with rear view of DB9 connector to PC

RS232 cable, meter to PC, P/N CBL01
Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.
Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.
Meter Setup Screens
Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.











Meter Setup Utilities




From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: List, Plot and Graph.
- List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.
- Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
- Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.


Laureate™ 1/8 DIN Case For Laureate Digital Panel Meters, Counters, Timers & Remote Displays

Key Features
- Meets 1/8 DIN Standard.
- Installs from front of panel.
- Short depth behind the panel: only 4" (102 mm) plus connectors.
- Understated 0.157" (4 mm) thick bezel.
- Meets NEMA 4X (IP-65) for high-pressure wawshdon when panel mounted.
- Screw clamps connectors meet VDE / IEC / UL / CSA safety standards.
- Rugged GE Lexan® housing material.
- Safety certified per EN 61010-1.
Dimensions

Maximum panel thickness: 4.5 mm (0.18")
Weight of base meter: 210 g (7.4 oz) typical (DPM, counter, timer, 6-digit remote display)
Weight of option boards: 30 g (1.0 oz) typical per board (analog output, relay output, communications)
Tightening Torque - Connectors: Screw terminal connectors: 5 lb-in (0.56 Nm)
Tightening Torque - Pawls: Digital Panel Meter Case Pawls: 5 lb-in (0.56 Nm)
Dimensioned CAD assembly drawings in EPRT, STEP, x_t. dwg, pdf file formats: Laureate-meter-case.zip (zipping prevents browser from opening CAD files as text files).
Panel Mounting
Slide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel.
The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter.
Turning each screw counterclockwise loosens the pawl and retracts it into its well. This position allows installed meter to be removed from their panel, or new meters to be installed in a panel. Do not remove the screws from their pawls. Doing so would cause the screw and pawl to fall off and likely get lost. Do not overtighten so as not to damage the plastic parts.
Process Signal and Ratiometric Meter Frequently Asked Technical Questions
Ratiometric potentiometer follower application
In this application shown, the signal from a sliding contact voltage divider can be converted to engineering units such as position, level or percentage. By operating in a ratiometric mode, the meter automatically removes any effects caused by variations in the excitation supply. Connect excitation return to signal return, and jumper the excitation output to 5 Vdc.
![]() ![]() |
![]() |
Powering two-wire transmittersThe isolated 24 Vdc, 50 mA excitation output, which is standard with all Laureate meters, is ideal for powering two-wire, 4-20 mA transmitters. The same two wires are used to apply voltage and carry the output current. Inside the meter, the 4-20 mA current is dropped across a 10 ohm resistor and sets up a 40-200 mV voltage, which is then sensed by the meter and scaled to engineering units.
|
Testing with peak detectionDestructive testing is an ideal application for the Laureate strain meter. Peak readings are automatically captured at rates up to 60 per second, while the display updates at a legible 3.5 readings per second. The peak reading can be recalled at the push of a button or be transmitted via RS-232 or RS-485. The meter provides isolated 10 Vdc power for up to four (4) the strain gauges and can be scaled to read out directly in engineering units from -99,999 to +99,999.
|
![]() |
![]() |
Custom curve linearizationThe Laureate DC meter with the Extended main board option allows exceptionally accurate custom curve linearization. For setup, up to 180 data points can be entered into a spreadsheet. The system then creates multiple non-linear spline-fit segments, which provide much better accuracy than linear segments. Illustrated, is the readout of volume of irregularly shaped tanks based on measured liquid level or pressure. Altimeters and thermistors are further applications.
|
Rate from successive readingsThe Extended computer board allows the display of rate based on successive readings, for instance flow rate based on changes in liquid level or static pressure in a tank. In the above illustration, the meter displays the rate in gallons at which a horizontal tank is being emptied. The input to the meter can be nonlinear, since only the linearized readings are compared for the determination of rate.
|
IPC
Splashproof Cover
CBL01
RS232 Cable for Meters
CBL02
USB-to-RS232 Adapter Cable
CBL04
RS232 Cable for LT Transmitters
CBL05
USB Data Cable for Meters
CBL06
USB-to-RS485 Adapter Cable
CBL07
USB Programming & Data Cable
CBL08
RS485 Splitter Cable
BKBD
RS485 RJ11 Terminal Block Adapter
Modular Design for Maximum Flexibility at Minimum Cost
All boards are isolated from meter and power grounds. Optional Plug-in-Play boards for communications and control include Ethernet, WiFi, serial communication boards, dual or quad relay boards, and an analog output board. Laureates may be powered from 85-264 Vac or optionally from 12-32 Vac or 10-48 Vdc. The display is available with bright red or green 0.56" (14.2mm) high LED digits. The 1/8 DIN case meets NEMA 4X (IP65) specifications from the front when panel mounted. Any setup functions and front panel keys can be locked out for simplified usage and security. A built-in 5, 10, 12, or 24 Vdc excitation supply can power transducers, eliminating the need for an external power supply. All power and signal connections are via UL / VDE / CSA rated screw clamp plugs.
The Laureate™ Series features modular design with up to 7 isolated plug-in boards, applicable to all Laureate 1/8 DIN Panel Meters.
Modular Hardware
The design of the Laureate™ Series is modular for maximum flexibility at minimum cost. All boards are isolated from meter and power grounds. The base configuration for panel meters or counter consists of a main module (with computer and plug-in display boards), a power supply board, and a signal conditioner board. Optional plug-in-play boards include an isolated setpoint controller board, an isolated analog output board, and an isolated digital interface board. Modular design and a choice of plug-in options allow the Laureate to be customized for a broad range of applications from simple monitoring to control and computer interface. There can be up to five plug-in boards in a 1/8 DIN Laureate.

Connecting Laureate Panel Meters to a Local Area Network (LAN)
Up to 30 Laureate Panel Meters and/or LT Transmitters can be configured for RS485 and daisy-chained to an LT Transmitter using Laurel’s High Speed Ethernet-to-RS485 converter board for seamless LAN integration. Alternatively, Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration.
Flexible Communication Options for Panel Meters
Laureate Panel Meters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.
What are 1/8 DIN Panel Meters for Process & Ratiometric Meter?
Panel Meters are essential tools in industrial and process control environments, where precise measurement and display of data are critical for operational efficiency and safety. Among these, 1/8 DIN Panel Meters are highly valued for their compact size, reliability, and versatile functionality. Specifically designed for process and ratiometric measurement applications, these Panel Meters provide accurate monitoring of variables such as temperature, pressure, flow, and signal ratios. This section explores the features, functionality, technical specifications, and diverse applications of 1/8 DIN Panel Meters in process and ratiometric measurement contexts.
Understanding the 1/8 DIN Standard
The term "DIN" stands for Deutsches Institut für Normung, the German Institute for Standardization, which establishes international standards for industrial components. The "1/8 DIN" designation refers to the size of these Panel Meters, which occupy one-eighth of a full DIN standard panel, with dimensions of approximately 96 x 48 mm (3.78 x 1.89 inches). This compact size allows 1/8 DIN Panel Meters to be seamlessly integrated into control panels, dashboards, or instrumentation setups, making them ideal for environments where space is limited but precise data display is essential.
What are Panel Meters?
Panel Meters are electronic devices designed to display digital or analog readings of various input signals, including voltage, current, temperature, pressure, flow, and more. They are widely used across industries for monitoring and controlling critical parameters. Equipped with high-visibility LED or LCD displays, these Panel Meters provide clear and precise readouts, reducing the potential for human error compared to traditional analog meters. Advanced features like programmable setpoints, alarms, and communication interfaces enhance their versatility, making them indispensable in modern industrial systems.
Process Meters: Monitoring and Control
Process meters, a subtype of Panel Meters, are specifically designed to measure and display process variables such as temperature, pressure, flow, level, or other environmental parameters. These Panel Meters are critical for maintaining and optimizing industrial processes, ensuring they operate within safe and efficient parameters. Key characteristics of process meters include:
-
Accuracy: 1/8 DIN Panel Meters provide highly accurate readings, essential for maintaining tight control over industrial processes where even minor deviations can lead to inefficiencies or safety risks.
-
Programmability: These Panel Meters offer configurable input types, allowing users to adapt them for various applications, such as switching between temperature and pressure measurements with a single device.
-
Alarms and Outputs: Equipped with programmable alarms, these Panel Meters can trigger alerts or control actions when process variables exceed predefined thresholds. They also support outputs like relays or analog signals for integration with other systems.
Ratiometric Meters: Precision in Measurement
Ratiometric meters, another subtype of Panel Meters, specialize in measuring and displaying the ratio between two input signals. This capability is particularly valuable in applications where relative changes are more critical than absolute values, such as in strain gauges, potentiometers, or sensors with ratiometric outputs. Key features of ratiometric meters include:
-
Application: Ratiometric Panel Meters are used in scenarios requiring comparison of two variables, such as monitoring the ratio of pressures between two points in a system to ensure balanced operation.
-
Stability: By focusing on the ratio of inputs, these Panel Meters minimize the impact of power supply fluctuations or environmental changes, ensuring stable and reliable measurements.
Combining Process and Ratiometric Measurement
In many industrial applications, there is a need to monitor both absolute process variables and relative signal ratios simultaneously. 1/8 DIN Panel Meters that support both process and ratiometric measurements offer exceptional flexibility. For example, they can measure the absolute temperature in a chemical reactor while also monitoring the ratio of temperatures between two points to ensure uniform heating. This dual functionality enhances precision, reduces equipment needs, and improves process efficiency and product quality.
Advantages of 1/8 DIN Panel Meters
-
Compact Size: The 1/8 DIN size allows these Panel Meters to fit into crowded control panels or enclosures, optimizing space without compromising functionality.
-
Versatility: Capable of handling both process and ratiometric measurements, these Panel Meters reduce the need for multiple instruments, streamlining system design.
-
Customization: Programmable display options, input types, and output configurations allow these Panel Meters to be tailored to specific application requirements.
-
Ease of Use: High-visibility digital displays provide clear, precise readouts, minimizing errors and simplifying operation for users.
-
Durability: Built with rugged enclosures and high IP ratings (e.g., IP65), these Panel Meters are designed to withstand harsh industrial environments, including dust, moisture, and vibration.
Technical Specifications
Typical specifications for 1/8 DIN Panel Meters for process and ratiometric measurements include:
- Input Types: Supports thermocouples, RTDs, pressure transducers, flow sensors, strain gauges, and potentiometers.
- Accuracy: ±0.01% of full scale or better.
- Display: 4-6 digit LED or LCD with customizable brightness and decimal points.
- Power Supply: 12-24V DC or 85-265V AC.
- Communication: RS-232, RS-485, Modbus RTU, or Ethernet.
- Environmental Rating: IP65 or higher for front panel protection.
- Operating Temperature: -10°C to 50°C, suitable for industrial environments.
Conclusion
1/8 DIN Panel Meters are compact, versatile, and highly functional instruments that play a pivotal role in process control and ratiometric measurement applications. Their ability to monitor absolute process variables and relative signal ratios with high accuracy makes them indispensable in industrial settings. With their space-efficient design, robust features, and seamless integration capabilities, these Panel Meters ensure precise control, operational efficiency, and safety across a wide range of industries.
Where are 1/8 DIN Panel Meters for Process & Ratiometric Meter Used?
Introduction
1/8 DIN Panel Meters are widely recognized for their compact size, precision, and versatility in industrial applications. Designed to fit into a standardized 1/8 DIN cutout (96 x 48 mm), these Panel Meters are ideal for monitoring and controlling critical parameters in space-constrained environments. When used for process and ratiometric measurements, they provide real-time data for variables like temperature, pressure, flow, and signal ratios, supporting efficient and safe operations. Below is a detailed exploration of the industries and applications where these Panel Meters are deployed.
What are 1/8 DIN Panel Meters?
1/8 DIN Panel Meters are display devices used to monitor real-time data from sensors, transducers, or other input devices in industrial processes. The DIN standard ensures compatibility with control panels, making these Panel Meters easy to install and integrate. They are equipped with high-resolution displays and advanced features like programmable inputs, alarms, and communication interfaces, making them suitable for both process monitoring and ratiometric measurement applications.
Applications of 1/8 DIN Panel Meters
-
Process Monitoring
- Temperature Control: In industries such as manufacturing, food processing, and HVAC, 1/8 DIN Panel Meters monitor temperatures from thermocouples or RTDs, ensuring optimal conditions in ovens, refrigeration units, or heating systems.
- Pressure Measurement: In chemical plants, water treatment facilities, or oil and gas operations, these Panel Meters display pressure readings from transducers, maintaining safe and efficient process conditions.
- Flow Rate Monitoring: Connected to flow sensors, these Panel Meters track the flow of liquids or gases in pipelines, supporting material handling, water treatment, or chemical dosing applications.
-
Ratiometric Measurement
- Proportional Control: In dosing or mixing systems, 1/8 DIN Panel Meters measure and display the ratio between two inputs, ensuring precise control over mixtures in chemical or pharmaceutical processes.
- Torque and Load Monitoring: In mechanical systems, these Panel Meters monitor torque or load relative to another parameter, such as motor speed, ensuring safe and efficient operation.
- Voltage and Current Ratios: In electrical systems, ratiometric Panel Meters compare voltage or current ratios for applications like power distribution or load balancing.
Industries Using 1/8 DIN Panel Meters
- Manufacturing: 1/8 DIN Panel Meters are used in automated manufacturing to monitor process variables like temperature, pressure, or flow, ensuring product quality and operational efficiency.
- Energy and Utilities: In power plants, water treatment facilities, or renewable energy systems, these Panel Meters monitor critical parameters to maintain system integrity and optimize energy use.
- Automotive and Aerospace: In these industries, ratiometric measurements ensure the safety and reliability of systems involving fuel, hydraulics, or environmental controls.
- Food and Beverage: Precise control of temperature, pressure, or flow is vital for maintaining product quality and meeting food safety standards, making these Panel Meters essential.
- Pharmaceuticals: In drug manufacturing, 1/8 DIN Panel Meters monitor process variables and ratios to ensure consistent formulation and compliance with regulatory standards.
- Chemical Processing: These Panel Meters track process parameters and signal ratios to maintain safe and efficient chemical reactions and production processes.
Conclusion
1/8 DIN Panel Meters for process and ratiometric measurement are critical tools across a broad spectrum of industries, including manufacturing, energy, automotive, aerospace, food and beverage, pharmaceuticals, and chemical processing. Their compact size, high accuracy, and versatile functionality make them ideal for monitoring and controlling critical parameters, ensuring operational efficiency, safety, and compliance. As industries continue to adopt advanced automation and data-driven technologies, these Panel Meters will remain essential for maintaining precise control and optimizing performance.
Less Information.